Display of Several Measuring Points
00: 23.12 °C 32.67 °C
Measured value including
minimum value:
MeasValList: Min value
00: 23.12 °C 19.34 °C
Measured value including
average value:
MeasValList: Avg. value
00: 23.12 °C 25.45 °C
Measured value including
limit value max:
MeasValList: Linit max
00: 23.12 °C 32.67 °C
Measured value including
limit value min:
MeasValList: Limit min
00: 23.12 °C 19.34 °C
Measuring range
only (again max. 20 channels):
MeasValList: Range
00: NTC °C
The programming functions can be accessed
as follows:
PROG
,
/
...
10.6 Assistant-Menus for Special Meas. Operations
Special measuring operations, i.e. thermal coefficient or wet bulb globe tem-
perature, require a series of sensors in a particular arrangement and function
channels programmed for calculating the required variables. To ensure that
these two special measuring operations can be performed quickly and easily
there is a special assistant menu for each.
10.6.1 Thermal Coefficient
To determine the thermal coefficient
q
/
T1
−
T0
the two temperature sensors
are connected as required (see manual, sec-
tion 3.2) to channels M0 and M1, and the
heat flow plate to M2. The temperature dif-
ference T(M1)-T(M0) is obtained automati-
cally on channel M9 with range
Diff
.
For this measuring operation the following
programming steps are needed :
Averaging mode on M9 :
CONT
or
CYCL
Averaging mode on M2:
CONT
or
CYCL
Range on M12:
q/dt
default reference channels: Mb1 = q = M02
Mb2 = Diff = M09
Enter the cycle by means of:
Cycle timer
Start measuring by means of:
<START>
Stop measuring by means of:
<STOP>
ALMEMO
®
2890-9
41
Assistant-Menu
Thermal Coefficient:
Inner temp. T0:
Channel: 00
00: 21.67°C NiCr
Outer temp. T1 : Channel: 01
01: 11.42°C NiCr
Difference dt: Channel: 05
05: 10.25°C Diff
Averaging mode: CONT
Heat flow q
Channel: 02
02: 103.6 ¤¥
Averaging mode: CONT
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
±
Thermal coeff. Channel: 12
12: 193. ¤mK q/dt
1 Range: q/dt
Cycle-timer:
00:30:00 Sn
START MANU PRINT ESC