11. Programming With Programming-Menus
and storage of all measured values is also possible.
The continuous storage and the continuous output of the measured values can
be activated with the conversion rate using the following two functions.
Function Conversion rate:
Input s. 9.5
Conv. rate: 10M/s
half-continuous measuring point scan (basic setting):
<OFF>
Cont:
-
continuous measuring point scan:
<ON>
Cont:
Ø
continuous storage off:
Storing:
-
activate continuous storage:
<ON>
Ø
continuous output off:
Output: -
activate continuous output:
<ON>
Ø
At a conversion rate of 50 measurements/sec. the following limita-
tions due to shortened analysation times are to be considered:
It should be noted, however, when selecting the measuring rate,
that the higher the measuring rate so the lower the measuring qual-
ity and, conversely, the lower the rate, the higher the quality.
At measuring rates above 10 mops (measuring operations per sec-
ond) mains hum suppression is not possible; as a result accuracy
may be adversely affected by interference over the connection
lines; (wherever possible use twisted wires).
At 100 or 400mops saving to memory is only possible with the mul-
timedia card but not with the internal EEPROM.
11.1.4 Start date and time-of-day, end date and time-of-day,
measuring duration
A sequence of measurements can, at certain points in time, be automatically
started and stopped. For this purpose the start time and date and the end time
and date can be programmed. If a date has not been specified the measure-
ment will be performed every day in the defined period. The current time must
be programmed beforehand. Or, alternatively, instead of specifying the end
time-of-day the measuring duration itself can be programmed.
Measuring duration function (format = hh:mm:ss)
Measuring duration 00:00:00
Function Start Time
(format
hh:mm:ss
):
Start time: 07:00:00
Function End Time
(format
hh:mm:ss
):
End time: ------
Function Start Date
(format
dd:mm:yy
):
Start date: 01.05.00
Function End Date
(format
dd:mm:yy
):
End date: ------
To clear values, select function and then use:
<OFF>
If the start time of a measurement has been pro-
grammed, the status line will show the symbol:
´
l©
´
If the end time of a measurement has been pro-
grammed, the status line will show the symbol:
´
©l
´
48 ALMEMO
®
2890-9