The locking and calibration menu (option KL)
11.8 The locking and calibration menu (option KL)
In the
"Locking and calibration" menu
you can
lock the right-of-access to certain menus and to
certain functions. Here you can also see the
serial numbers and calibration data for the de-
vice itself and for any sensors attached. With
option KL it is possible not only to correct the
sensor at several points in the connector itself
but also to manage the associated calibration
data.
The right-of-access to this and other menus and
to certain key functions can be stipulated in de-
tail and protected by password by means of the
parameters "Menu" and "Fct".
Menu Locking the menus
0
none
1
+ calibration menu, except password
2
+ programming menus, except recording to memory and output from memory
3
+ recording to memory and output from memory
4
+ wizard menus
5
+ measuring menus, except user menu U1
Fct Locking the functions
and keys
0
none
1
+ data input, switching on and off
PROG, ON, OFF, ZERO, ADJ
2
+ clearing measured data
CMEM, CLR, CLRA
3
+ start / stop / output measuring operation
START/STOP, MANU, ARRAY, PRINT
4
+ function selection, measuring point selection
PROG, F«, M«
No password, locking with new password:
Password :
-
-
-
-
Locked with password, enter correct password :
Password :
****
Select locking level, menu, and function :
Locking mode : Menu: 0 Fct: 0
The device type (with version and serial number) and the sensors (with order
number and serial number) are displayed. With option KL you can enter the
date of the next calibration and the calibration interval in months. If “Calibration
message” is activated then, as soon as the next calibration is due, a message
to this effect will appear when the device is switched on.
ALMEMO
®
2890-9
73
* LOCKING-CALIBRATION *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Password:
****
Locking level: Menu: 0 Fct: 0
Device:
2890-9 6.22
Serial number:
04020123
Next calibration:
01.12.05
Signal for calibration:
Ø
Sensor:
Channel: 00
Type:
FHA646-6
Serial number:
04020123
Next calibration::
01.02.06
Calibr. interval:
12 Month
PRINT ESC