10. Measuring using Measuring-Menus
10.4.2 Averaging Mode
A detailed description of averaging over measuring point scans is given in the
manual section 6.7.4. The type of averaging is determined through the function
Averaging Mode
. With the averaging mode and the corresponding operation
the following modes can be set:
Function no averaging:
Averaging Mode: -----
Averaging over single measurements with MANU or
all measured values from START to STOP:
CONT
Averaging over all measured values of a cycle:
CYCL
Lightening up for control purposes if averaging is in progress:
»
Display of the average value in the function
Average value 12.34 m/s
For recording average values a function channel with the range M(t) or the corre-
sponding output function M(t) is required instead of the measured value. (s. 10.5.5).
10.4.3 Averaging over Manual Single Measurements
Single manual meas. point scans E
i
are performed for an averaging of isolated
measurements at certain points or times. At all measuring points where mea-
sured values must be averaged the averaging must be switched on by using
the averaging mode ´
CONT
´.
1. Stop measurement, if it is started:
<STOP>
2. Set the averaging mode (s. 9.5):
Averaging mode: CONT
For smoothing a meas. value select damping:
Damping: 20
Switch the continuous scanning off, if required:
Conv. rate:10M/sCont:
-
3. Clear avg. value by selecting it (s. 9.4) and using:
<CLR>
Function ´Average value´ displays:
Average val: ----- m¡
Function ´Counts´ displays:
Counts: 00000
4. Manual scans Ex of single measurements:
<MANU>
Function ´Average value´ displays:
Average val: 12.34 m¡
Function ´Counts´ displays:
Counts: 00001
5. Repeat step 4 for each measuring point.
6. Output of all function values of the menu by using:
<PRINT>
34 ALMEMO
®
2890-9
E1
E2
E3
E4
MANU
MANU
MANU
MANU
CLR
M
=
∑
i
E
i
/
N