11. Programming With Programming-Menus
11.5.7 Hysteresis
In case of an exceeding of limit values the hysteresis of an alarm condition
can, generally for all sensors, be set in the function
Hysteresis
(s. 11.3.5
and manual 6.2.7) within the range from 0 to 99 digits (standard setting is 10
digits).
Change hysteresis (0 to 99) s. 9.5:
Hysteresis: 10
11.5.8 Operating Parameters
Some operating parameters are software options and can be user-configured
with the function
Configuration
(s. manual 6.10.13.2).
Mains frequency noise suppression 60Hz instead of 50Hz
Configuration: F-------
Clearing all measured values on start of a measurement
Configuration: -C------
Ring memory (overwriting of old data when full)
Configuration: --R-----
Switching off the signal transmitter
Configuration: -----S--
The following parameters can be used to check the device function:
Out of 20 channels 5 are activated:
Meas.Chan:20 active:05
Cold junction temperature = socket temperature:
CJ-Temperature: 25.4°C
11.6 Output Modules
The data logger ALMEMO
®
2890-9 has two
output sockets, A1 and A2, to allow an output
of the measured values as analog or digital
values or as alarm signals. Furthermore, it is
possible to trigger various functions through
trigger pulses. To suit all possibilities, while
keeping the hardware requirements at a mini-
mum, all necessary interfaces have been in-
tegrated into the ALMEMO
®
output con-
nectors.
The output modules, like the sensors, are au-
tomatically detected and displayed in the
menu
OUTPUT MODULES
.
With the relay trigger analog modules certain function variants can be config-
ured (see 11.6.2Fehler: Referenz nicht gefunden), relays can have certain limit
values assigned to them (see Fehler: Referenz nicht gefunden), or analog out-
puts can be assigned to certain measuring channels. In this menu all ports can
be selected and configured accordingly. The connection possibilities are de-
scribed in the instructions for the output module concerned.
68 ALMEMO
®
2890-9
* OUTPUT MODULES *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Output socket A1:
DK Data cable
0: RS232
Baudrate: 57.6 kBd
Output socket A2:
EA Trigger-Alarm
2: Rx int. assigned
Relay: 01------
Analogue channel: 00
Analogue value: +32500
PRINT ESC