Output Modules
11.6.1 Data cable
Via the serial interface you can output cyclic data logs, all the function values
from the measuring menus, and all the programming details for the device and
for the sensors to a printer or computer. The ALMEMO
®
data cables and their
connection to devices are described in detail in the Manual, Section 5.2. Other
modules for networking the devices are described in detail in the Manual, Sec-
tion 5.3. All available interface modules can be connected to socket A1
(2)
; this
is with the exception of cable ZA 1999-NK, which is used for networking extra
devices; this must be connected to socket A2.
Under the prevailing socket the menu indicates:
Outout Socket A1:
DK Data cable
Variant 0: Standard interface always active
0: RS232
The baud rate is also stored in the cable connector:
Baudrate: 9600 Bd
11.6.2 Relais trigger modules
Whereas, for the purposes of addressing peripheral devices via the relay and
trigger input (see Manual 5.1.2/3), V5 modules (ZA 1000-EAK, ZA 8000-RTA)
provide only one function variant (see Manual
6.6.4), the new V6 relay trigger analog adapter
ZA 8006-RTA3 offers up to 10 relays or op-
tions with 2 of these as trigger inputs or up to 4
as analog outputs. All elements can be config-
ured individually for these functional variants.
Old output cables can be reconfigured for V6
functions by means of AMR-Control. These
modules can be connected equally well to out-
put socket A2 or output socket A1
(2)
. To en-
sure that all elements are addressed, each of
these sockets has been assigned 10 port ad-
dresses.
Socket Connection
Port addresses
A1
V6 output modules at socket A1
10..19
A2
V6 output modules at socket A2
20..29
In the
"Output modules "
menu the elements of the output modules can be in-
dividually selected and functions programmed as follows (see Manual 6.10.9) :
First
select the port
by pressing :
<P>
:
▲
or
▼
e.g. port 0, socket A2 (port address 20) :
Port: 20
The element concerned is recognized :
Relay :
Relay type = NO (normally open) :
Relay : NO
Relay type = NC (normally closed) :
Relay : NC
Relay type = changeover :
Relay : Changeover
ALMEMO
®
2890-9
69
* OUTPU MODULES *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
ZA 8006-RTA3 Socket: A2
Port: 0
Adr.: 20
Relais: normally open 0.5A
2: Relais driven externally
Status: activ open
Watchdog: Ø Error
ON OFF P PRINT ESC