CLEANING
A useful guide to decontamination is MHRA leaflet DB2003(06).
Do not use abrasive cleaners as this could damage the surface of the equipment. For
cleaning, wipe with a clean cloth soaked in detergent solution and wring out. Depending
on the detergent manufacturer’s instructions the solution may be heated.
Wipe all surfaces of the equipment and the cloth should be re-immersed periodically and
the detergent changed frequently. The detergent should be rinsed or wiped off using a
clean cloth immersed in clean water and wrung out.
To disinfect make sure the equipment has been cleaned and dried and then wipe over with
a clean cloth and a compatible liquid disinfectant. After disinfecting clean with clean water
and dry.
When using detergent or disinfectants then please refer to manufacturers’ instructions and
if unsure then consult your local Infection Control Specialist for advice.
This equipment can be cleaned/disinfected by moist heat at 80oC for a minimum of 1
minute or autoclaved at 80oC for at least 1 minute. After cleaning or disinfecting treat all
joints with a water dispersing agent, e.g. WD40.
REISSUE
If you reissue or are about to reissue this product, please thoroughly check all components
for their safety.
This includes:
• Tightness of all nuts/bolts/castors and other screw in/bolt in/push in components
• Upholstery (if applicable) for security, splits, exposed staples, etc
• Straightness and integrity of all tubing and welds
If in any doubt, please do not issue or use, but immediately contact your supplier for
service support.
When reissuing, a risk assessment must be undertaken as to the suitability of the Bed
Transfer Aid for the particular user and user’s bed.
IMPORTANT INFORMATION
The Bed Transfer Aid hand grips have been designed to meet the latest MHRA guidelines
on entrapment, however you must ensure that there is no risk to the user or carer after the
installation and service of the Bed Transfer Aid with Strap or without Strap.
Do not allow children or unauthorised persons to play with the equipment or use the bed
transfer aid without proper supervision.
CARE & MAINTENANCE
Please make a safety check of the product at regular intervals or if you have any concerns.