14
The SYNCHRONOUS ON/OFF key is in addition to the individual
OUTPUT switches and permit both outputs to be turned on or off
synchronously with a single key press. Since this key turns both outputs
on or off with alternate presses it is necessary for both outputs to be in the
same state (i.e. both on or both off) before the key is used. Pressing the
key with one output on and one output off will have no effect.
The V
TRACK
status at power on is the same as at last power off.
Note that for CPX200D an OVP (over-voltage protection) trip on one output in Voltage Tracking
mode will only trip that output off, leaving the other output on; this is also the default mode for
CPX200DP.
However, for CPX200DP, it is possible to configure both the OVP and OCP (over-current
protection) trip by remote commands such that both outputs are switched off when either output
experiences an OVP or OCP trip; see Remote Commands section for further details. Once set, this
configuration applies in both remote and local operating modes until changed by remote command,
or reset to default mode by *RST or by turning the POWER switch to standby then on again.
Synchronous Output On/Off Control