INTERFACES
Full digital remote control facilities are available through the RS232, LAN and USB interfaces.
Setting and readback resolutions are the same as the Output and Meter specifications
respectively.
RS232:
Standard 9-pin D-connector. Baud rate 9600.
USB:
Standard USB 2.0 hardware connection.
LAN:
Ethernet 100/10base-T hardware connection.
LXI V1.2, Class C compliant.
GPIB (factory option):
Conforming with IEEE488.1 and IEEE488.2.
Remote Command
Processing Time:
Typically <100ms between receiving the command terminator for a
step voltage change at the instrument and the output voltage
beginning to change.
GENERAL
AC Input:
110V – 240V AC ± 10%, 50/60Hz.
Installation Category II.
Power Consumption:
1600VA max.
Operating Range:
+5ºC to +40ºC, 20% to 80% RH.
Storage Range:
−
40ºC to + 70ºC.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Store/Recall:
Up to 10 set-ups can be saved and recalled via the keyboard or
remote interfaces.
Safety:
Complies with EN61010-1.
EMC:
Complies with EN61326.
Size:
130mm H (3U) x 356mm W x 413mm D
Weight: 9.2kg
Options:
19-inch rack kit.
4