2
Specification
General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications
apply for the temperature range 18°C to 28°C after 1 hour warm-up with no load and calibration
at 23°C. Typical specifications are determined by design and are not guaranteed.
OUTPUT SPECIFICATIONS
Voltage Range:
0V to 60V
Current Range:
0.01A to 50A
Power Range:
Up to 1200W
Voltage Setting:
Resolution 1mV
Accuracy: 0.1% of setting ± 2mV
Current Setting:
Resolution 10mA
Accuracy: 0.3% of setting ± 20mA
Operating Mode:
Constant voltage or constant current with automatic cross-over
provided that the power demanded stays within the power envelope,
see graph. Outside of this envelope the output becomes unregulated.
CV, CC or UNREG mode indication in display.
Output Switch:
Electronic, with ON indication.
Preset voltage and current displayed when off.
Output Terminals:
Front panel safety terminals accepting 6mm wire diameter, 6mm
plugs or 8mm spades at 50 Amps max, or 4mm plugs at 30 Amps
max. Duplicate rear panel terminals.
Sensing:
Switchable between local and remote. Screwless remote sense
terminals on both front and rear panels. Sense miswiring trip and
indication.
Ripple & Noise
(20MHz bandwidth):
Typically <3mVrms, <20mV pk-pk at maximum load, CV mode
Load Regulation:
Change in output for any load change within PowerFlex envelope,
using remote sense:
Constant voltage: <0.01% ± 5mV
Line Regulation:
Change in output for a 10% line change:
Constant voltage: <0.01% ± 5mV
Transient Response:
<250us to within 100mV of set level for a 5% to 95% load change.
QPX 1200