Monnal T75 user manual
Note:
The emission characteristics of this device allow it to be used in industrial areas and
in a hospital environment (class A as defined in CISPR 11). If used in a domestic environ-
ment (for which class B as defined in CISPR 11 is normally required), this equipment might
not offer adequate protection to radio-frequency communication services. The user may be
required to take mitigation measures, such as relocating or re-orienting the ventilator.
CAUTION:
Exceptional electromagnetic disturbance may degrade ventilation performance.
CAUTION:
RF hand-held communication devices should not be used (including peripherals
such as antenna cables and external cables) within a radius of 30 cm (12 inches) of any part
of the
Monnal T75
, including the cables specified by the manufacturer. Otherwise, the perfor-
mance of this equipment may be impaired.
Length of cables
CAUTION:
These accessories and cables must be used with the
Monnal T75
. Their use with
other medical electrical equipment may increase the emissions of the equipment or decrease
its immunity.
Accessories and ca-
bles used during test-
ing
Max-
imal
length
Test type
in compliance with
RF Emission
CISPR 11, Class B
Emission of harmonic currents
IEC 61000-3-2
Voltage fluctuation and voltage flickers IEC 61000-3-3
Electrostatic discharge immunity
IEC 61000-4-2
Radiated immunity – Electromagnetic
fields
IEC 61000-4-3
Electrical fast transient/burst immunity IEC 61000-4-4
Surge immunity
IEC 61000-4-5
Immunity to conducted disturbances,
induced by radio-frequency fields
IEC 61000-4-6
Radiated immunity - Magnetic fields
IEC 61000-4-8
Monnal T75
equip-
ment:
Power cable (3 m) -
KY120500
3 m
Immunity to voltage dips, short inter-
ruptions and voltage variations
IEC 61000-4-11
Recommended separation distances
The medical device is intended for use in an electromagnetic environment in which the disturbances
due to RF radiation are controlled.
The user or installer of the medical device can help prevent electromagnetic interference by maintain-
ing a minimum distance, according to the maximum output power of the radio-frequency transmis-
sion equipment. RF hand-held communication devices should not be used (including peripherals such
as antenna cables and external cables) within a radius of 30 cm (12 inches) of any part of the
Monnal
T75
, including the cables specified by the manufacturer. Otherwise, the performance of this equipment
may be impaired.
114
Summary of Contents for MONNAL T75
Page 2: ......