Monnal T75 user manual
used exceeds the radio-frequency compliance level specified above, the product performance must be
tested in order to check whether or not it is compliant with the specifications. In the case of any abnor-
mal performance of the device, additional measures may be necessary, such as reorienting or relocat-
ing the product.
12.4.3.4 Pneumatic specifications
High pressure (HP) and low pressure (LP) O2 inlets
Type of gas fitting
NF, DISS, NIST (HP)
Connector (LP)
O2 pneumatic supply
2.8 - 6 bar / 280 - 600 kPa / 40 - 86 psi (HP)
0 – 1.5 bar / 0 – 150 kPa / 0 – 22 psi (LP)
Required maximum flow rate (at atmospheric
pressure)
105 L/min at 2.8 bar / 130 L/min at 6 bar (HP)
85 L/min at 1.5 bar (LP)
Standalone mode
Cylinder (HP)
Concentrator (LP)
Mixer
Electronic, electrochemical O2 sensor
Precision (% of set value)
< 3%
Gas consumption(1)
Patient venti 4 L/min. (flow-by + internal
consumption)
Connectors
Inspiratory hose connector
ISO 22 mm male
Expiratory hose connector
ISO 22 mm male
Inspiratory and expiratory resistances(2)
Resistances at 60 L/min (cmH2O)
(Unit + single-use adult patient circuit KG019300)
Inspiratory: 3.9
Expiratory: 3.2
Resistances at 30 L/min (cmH2O)
(Unit + single-use paediatric patient circuit
KG019400)
Inspiratory: 5.7
Expiratory: 4.9
Pressure
Maximum limited pressure
(P lim max)
100 cmH2O; blower performance limitation
Maximum working pressure
(P w max)
0-100 cmH2O
118