Monnal T75 user manual
Curves
Loop curves
Pressure / Volume
Volume / Flow rate
Flow rate / Pressure
CO2/Volume (only for KB033600)
Monitoring
Conditions for measurement of flows and vol-
umes
BTPS
Data storage
Trends
Simultaneous display of two parameters mea-
sured during a maximum period of 80 h.
All measured parameters are accessible.
Alarm log
List of alarms triggered during the use of the unit
(4000 events recorded (1))
(1) The unit records 4000 events. An event is represented by a set-point or an alarm threshold when an
adjustment is made and an alarm if triggered. Because the alarm log reports only alarm-related events,
the maximum number of events recorded can be below 4000.
12.4.3.7 Measured parameters
Measured para-
meters
Measurement
range
Resolution
Precision
Filtering
Peak airway pres-
sure (Ppeak,
cmH2O)
0 - 100
1
+/- (2 cmH2O +
4%)
15 ms
Positive end of ex-
piratory pressure
(PEEP, cmH2O)
0 - 99
1
+/- (2 cmH2O +
4%)
15 ms
Plateau pressure
(Pplat, cmH2O)
2 - 99
1
+/- (2 cmH2O +
4%)
15 ms
Frequency (RR,
bpm)
0 - 120
1
+/-1 bpm
4 cycles
Insufflated tidal
volume (VTi, mL)
0 - 3000
1
+/- (4 mL + 15%)
-
Leak (%)
0 - 100
1
+/- 20%
-
Expired tidal vol-
ume (VTe, mL)
0 - 5000
1
+/- (20 mL + 15%) -
Expired volume
per minute (MVe,
L/min)
0 - 40
0.1
+/- (0.5 L/min +
15%)
30 s
122