12 Technical description
Measured para-
meters
Measurement
range
Resolution
Precision
Filtering
Slope of the vol-
umetric capno-
gram on the
alveolar plateau
(SlopeCO2,%CO2/
L)
0 – 99
0,1
-
-
1) Filtering is performed with a rolling average of n samples.
Oxygen sensor specifications
Service life
Approximately 5000 h (variable duration as a
function of concentration and temperature)
Conditions of use and storage
Same as those of the device
Power supply
Provided by the device, including during battery
operation
Calibration
Calibration is automatically conducted during in-
teractive testing.
Precision is guaranteed only when there is fre-
quent calibration
(at least weekly).
Minimum flow rate to guarantee precision.
5 L/min
Drift of measurement accuracy over 6 hours
< +/-3 Vol.%
Response time at 90% (extreme conditions)
45s
Response time at 90% according to the EN ISO
80601-2-55 standard
< 10 s
Start time
Immediate
Sampling rate of the O2 cell data
5 ms
Influence of humidity on the oxygen measure-
ment
- 0.03 (% per%RH at 25°C)
Effect of pressure
Compensation of the measurement as a function
of atmospheric pressure.
Compensation of the measurement at average
pressure of the respiratory cycle.
12.4.3.8 Specifications of accessories
Single-use adult patient circuit
Part No.
KG019300
125