Monnal T75 user manual
Adult
Child
Infant
Alarms
Unit
Min.
Max.
Incr.
Default
Min.
Max.
Incr.
Default
Min.
Max.
Incr.
Default
T ap-
nea
s
15
60
1
20
4
60
1
20
2
30
1
10
CO2 alarms (if the CO2 monitoring software option is enabled)
mmHg
OFF, 1
98
1
27
OFF, 1
98
1
30
OFF, 1
98
1
30
etCO2
low low
%
OFF, 1
14
1
4
OFF, 1
14
1
4
OFF, 1
14
1
4
mmHg
1
99
1
49
1
99
1
49
1
99
1
49
etCO2
haut
%
1
15
1
6
1
15
1
6
1
15
1
6
VM-
CO2
high
mL/
min
OFF
990
10
OFF
OFF
990
10
OFF
OFF
990
10
OFF
VM-
CO2
high
mL/
min
10
1000
10
1000
10
1000
10
1000
10
1000
10
1000
VMalv
low
L/min
OFF
39
0,5
OFF
OFF
39
0,5
OFF
OFF
39
0,5
OFF
VMalv
high
L/min
1
40
0,5
40
1
40
0,5
40
1
40
0,5
40
Settings: PRVC and PS-PRO (Invasive ventilation)
Adult
Child
Infant
De-
scrip-
tion
Min.
Max.
Incr.
De-
fault
Min.
Max.
Incr.
De-
fault
Min.
Max.
Incr.
De-
fault
Maxi
Pas
Incr.
Expi-
ratory
pres-
sure
PRVC
PS
PRO
PEP
cmH2O 0
50
1
0
0
50
1
0
0
50
1
0
FIO2
PRVC
PS
PRO
FiO2
%
21
100
1
50
21
100
1
50
21
100
1
35
Fre-
quen-
cy
PRVC
f
c/min
4
80
1
15
5
120
1
25
5
120
1
40
VCRP
PI
AUTO
2
AUTO
99
AUTO
2
AUTO
99
AUTO
2
AUTO
99
Insuf-
fla-
tion
pres-
sure
PS
PRO
AI
cmH2O
AUTO
2
AUTO
40
1
AUTO
15
AUTO
2
AUTO
40
1
AUTO
15
AUTO
2
AUTO
40
1
AUTO
15
Pres-
sure
In-
crease
slope
PRVC
PS
PRO
Pente
cmH2O/
s
AUTO
AUTO
10
150
20
200
10
150
20
200
10
150
I:E
ratio
VCRP
Ti/
Ttot
%
2
40
1
33
5
60
1
33
5
60
1
33
128