Monnal T75 user manual
27
28
•
Connect the expiratory limb of the patient circuit to the expiratory valve of the ventilator: (27).
•
Connect the inspiratory limb of the patient circuit to the inspiratory outlet cone of the ventilator:
(28).
CAUTION:
Make sure to limit dead space while installing patient circuit and accessories.
Air Liquide Medical Systems
recommends the use of a bacteriological filter at the inspiratory outlet
of the ventilator or, preferably, at the Y-piece. It is also recommended to use patient circuits equipped
with water traps when using a humidifier.
2
1
1
Y-piece
2
Double-limb patient circuit
CAUTION:
Empty the water traps regularly during ventilation.
When using the equipment on a patient for the first time, make sure that you follow the hospital’s hy-
giene protocol for new single-use equipment or adequately sterilized reusable equipment. The acces-
sories and consumables (patient circuit, masks, expiratory valves, adapters, nebulizer, etc.) are gener-
ally available in single-use and autoclavable versions.
CAUTION:
Reusing single-use accessories or consumables carries the risk of patient cross-
contamination.
3.2.4 CO2 Measurement Probe
See CO2 Option on page 43 (for ref. KB033600).
3.2.5 Humidifier
If the ventilator is used with a humidifier ensure that it is always placed lower than the ventilator and
the patient.
Empty the water traps regularly in order to limit condensation in the pipes.