In the event of absence of patient respiratory activity, this mode provides the guarantee of controlled
ventilation (VCV).
When the patient has regular and detectable respiratory activity, the unit responds by:
•
supplying a “spontaneous” cycle (PSV) if the time elapsed since the last “controlled” cycle is less
than the SIMV period.
•
supplying a “controlled” cycle (VCV) if the time elapsed since the last “controlled” cycle is greater
than the SIMV period. If patient respiratory activity declines again after a “spontaneous” cycle, the
unit waits for the SIMV period - set TiMax to expire before triggering a “controlled cycle” (VCV) it-
self.
Insufflation in a spontaneous cycle with pressure support ends:
•
If the flow rate falls below the set expiratory threshold (E. Trig);
•
If the patient makes an expiratory effort,
•
or if the insufflation time reaches the maximum Ti setting (Timax).
Note:
Apnea ventilation can be activated. The apnea parameters should be suited to the con-
dition and requirements of the patient.
4.7.5 PSIMV (Synchronized Intermittent Mandatory Pressure Monitored Ventilation)
Principle
PSIMV Mode combines mandatory assisted pressure-controlled ventilation and spontaneous patient
ventilation between the assisted controlled cycles.
Ventilation set-points
FiO2
inspired oxygen fraction
PI
inspiratory pressure (cmH2O)
RR SIMV determines the frequency of the imposed cycles
(c/min)
PEEP
positive end of expiration pressure (cmH2O)
Tins
inspiration time of intermittent controlled assisted cycles (s)
I.Trig
inspiratory trigger (L/min or cmH2O)
PS
pressure support ventilation delivered during the spontaneous cycles (cmH2O)
TImax
maximum inspiration time of spontaneous cycles (s)
Slope
pressure support pressure rise slope (cmH2O/s)
E.Trig
expiratory trigger (% of peak inspiratory flow).
Operation
In this ventilation mode, the “controlled” cycle (PCV) delivers a fixed pressure at the set frequency
RRSIMV. A SIMV period arises from this frequency.
E.g. for a SIMV frequency set to 10 cycles per minute, the SIMV period between two “controlled” (PCV)
cycles is 6 seconds.
36