8 Alarms
N°
Alarm
Priority
Alarm activation
Alarm onset delay
Actions
79
High VTi!
LOW
Insufflated volume >
set threshold
Immediate triggering
The machine goes
into expiration mode
when the threshold
is reached.
Check the suitability
of alarm levels with
ventilation settings
In pressure-control
mode, check leakage
level
The measured fre-
quency is above the
threshold
Immediate triggering
Check the suitability
of alarm levels with
ventilation settings
80
High respiratory rate!
LOW
Self-triggering of in-
spiratory trigger
Immediate triggering
Lower the inspiratory
trigger sensitivity
81
-
86
Non Critical error! Ventilation Effec-
tive
LOW
Electronics failure
Immediate triggering
Use a different unit
and contact the tech-
nical department
87
Ventilator operates from internal bat-
tery!
LOW
[acknowledgeable]
Loss of mains power
supply
Immediate triggering
Battery life (see Bat-
page 99)
Connect the unit to
the mains if internal
power is not neces-
sary
88
Ventilator operates from external bat-
tery!
LOW [acknowledge-
able]
Loss of mains power
supply
Immediate triggering
Monitor the remain-
ing charge in the ex-
ternal battery and
connect the unit to
the mains if the bat-
tery is low
89
FiO2 sensor to be soon replaced!
LOW
Voltage below a
threshold
Immediate triggering
Replace the O2 sen-
sor the next time you
put the unit in ser-
vice.
90
Perform a battery test! (Machine
Configuration menu)
LOW
[acknowledgeable]
Every 6 months
Conduct battery test
91
Did you start the automatic tests with
the nebulizer?
LOW
[acknowledgeable]
Nebulization activat-
ed without having
run automatic test
with nebulizer con-
nected
Immediate triggering
Run the automatic
test
92
End-inspiratory occlusion in progress
INFO
“Inspiratory pause”
respiratory function
selected
Triggering at the
start of the expirato-
ry phase
93
End-expiratory occlusion in progress
INFO
“Expiratory pause”
respiratory function
selected
Triggering at the
start of the inspirato-
ry phase
96
Low pressure O2 activated
INFO
Press the “O2 low
presure” button
Immediate triggering
97
Nebulization in progress
INFO
Nebulization in
progress
Immediate triggering
87