8 Alarms
N°
Alarm
Priority
Alarm activation
Alarm onset delay
Actions
113
Intubation tube compensation en-
abled
INFO
Press the 'TC' button
Immediate triggering
114
Operate on altered O2 supply
INFO
'Enabled' setting of
altered O2 network
in administrator con-
figuration
Immediate triggering
115
NIF measurement in progress
INFO
Press the NIF key
Immediate triggering
116
Low pressure!!!
HIGH
Pressure in pa-
tient circuit < alarm
threshold
Triggered after 3
consecutive ventila-
tion cycles
Check the level of
leakage
117
Low pressure!
LOW
Pressure in pa-
tient circuit < alarm
threshold
Immediate triggering
Check the level of
leakage
118
Manual cycle in progress
INFO
Press the
Cycle.M key
Immediate triggering
119
CO2 apnea!!!
HIGH
CO2 sensor: No pa-
tient breaths have
been detected for
more than 20 sec-
onds
Immediate triggering
A clinician must re-
evaluate the patient's
ventilation
120
Check IRMA probe (CO2) adapter!
LOW
CO2 sensor: UU
adapter is not prop-
erly placed on the
probe
Immediate triggering
Check the UU
adapter
121
Replace IRMA probe (CO2) adapter!
LOW
CO2 sensor: UU
adapter is obstruct-
ed or blocked
Immediate triggering
Change the adapter
122
CO2 concentration is outside toler-
ance!
LOW
CO2 measured >
15%
Immediate triggering
If CO2 is correctly
< 15%, calibrate the
probe
123
Calibrate the IRMA probe (CO2)!
LOW
CO2 sensor: Probe
drift
Immediate triggering
Calibrate the probe
124
IRMA probe error (CO2)!
LOW
CO2 sensor: probe
hardware or soft-
ware error
Immediate triggering
Disconnect and re-
connect the probe. If
the problem persists,
change probes.
125
CO2 measurement is inoperative!
LOW
CO2 sensor: Probe is
not connected
Immediate triggering
Connect the IRMA
probe
126
IRMA probe (CO2): Internal tempera-
ture outside tolerance!
LOW
CO2 sensor: The in-
ternal probe temper-
ature exceeds the
maximum measure-
ment threshold
Immediate triggering
If the ambient tem-
perature is normal,
change probes
127
IRMA probe (CO2 Ambient pressure
is out of tolerance!
LOW
CO2 sensor: The at-
mospheric pressure
measured by the
probe exceeds the
maximum measure-
ment threshold
Immediate triggering
If the ambient pres-
sure is normal,
change probes
89