Monnal T75 user manual
100% The battery is fully operational
75%
75% battery life remaining, does not need to be replaced.
50%
50% battery life remaining.
The conditions of use of the battery must be taken into consideration when deciding
whether to replace it or not:
If the battery is never used apart from in the test (permanently connected to the AC power
supply, back-up supply), the remaining battery life may be sufficient.
If the battery is used often, during intra-hospital transfer of variable duration, it may be nec-
essary to replace it.
25%
The battery has reached its operating limit.
Please replace it.
In order to preserve a record of the test performed, the information is permanently displayed next to
the
[Battery test]
access key, and includes remaining battery life and test date.
Note:
This test resizes the battery charge indicator displayed on-screen: for example, if the
battery is down to 50% of its initial charge, no more than three of the five pictogram cells can
be lit. This function keeps the user aware of the overall state of the battery.
11.2 By the technician
Servicing once a year and checking the operation and performance of the unit.
The ventilator counts two operational parameters: number of hours of ventilation, and number of
hours that the machine is powered up.
These times can be viewed at any time in the stand-by screen.
The internal battery must be replaced after 500 charge/discharge cycles or 3 years (whichever comes
first), and this replacement must be performed by a qualified technician. For more information, refer to
the maintenance manual.
11.3 O2 cell
Note:
The working life of the O2 cell is approximately 5000 hours, but this can vary according
to the oxygen concentration used and the ambient temperature.
100