Monnal T75 user manual
Temperature
+10°C to +40°C (+50°F to +104°F) (See warning under table)
Relative humidity
0 to 90% without condensation at 40°C max.
Atmospheric pressure
700 to 1060 hPa
Storage and transport conditions
Temperature
-20°C to +70°C (-4°F to +158°F)
Relative humidity
0 à 90% without condensation at 40°C max
Atmospheric pressure
500 to 1060 hPa
Turbine and screen
Turbine warranty
40000h or 5 years (whichever comes first)
Control
Resistive tactile pad
Screen
Color 10.4", 640*480 pixels
Protection
Protection index
IP3X (if the serial number is less than MT75-05000)
IP31(if the serial number is greater than MT75-05000)
Applied parts according to IEC 60601-1
Applied parts
The respiratory system of the
Monnal T75
ventilator is a type B ap-
plied part.
The CO2 probe of the
Monnal T75
is a type BF applied part
CAUTION:
Ambient temperature warning: the temperature of the
Monnal T75
delivery flow
can exceed that of the room. Exercise extreme caution when using the
Monnal T75
at ambi-
ent temperatures above 38 C (100.4°F).
At an ambient temperature of 38 C or higher, gas temperature can climb as high as 43 C, de-
pending on the settings. The 43°C limit can never be exceeded. In such conditions, humidifi-
cation or nebulization may not be effective. When ambient temperature exceeds 38°C, Air Liq-
uide Medical Systems recommends use of HME filters.
12.4.3.2 Electrical specifications
Main power supply
Input voltage
100 - 240 V AC +/- 5%
Frequency
50 to 60 Hz
Electrical power consumption
250 VA
250 VA
T 16 A
External power supply fuse (x1)
T 3,15 A, 250 V
T 16 A
0,25 A FST
112