Monnal T75 user manual
Note:
The set periodicity remains in the memory when the function is disabled.
5.4 P0.1 measurement key
P0.1 measurement can be useful to determine whether extubation will be successful. The procedure
begins with detection of an inspiratory demand by creating end-expiratory occlusion and by measuring
the drop in pressure after 0.1 seconds. The inspiratory cycle then continues normally.
The measurement result is shown on the monitoring pad and in the trends, and in a window which is
displayed at the end of the selective measurement.
To begin occlusion pressure measurement:
•
Press the
[Measures]
button;
•
A new window opens with the
[R&C stat]
and
[P0.1]
’ functions.
The occlusion pressure measurement can be performed automatically when required (--) or according
to the following frequencies (min): 1, 2, 3, 4, 5, 10, 20, and 30.
1.
Press on the
[P0.1]
set-point on the bottom line;
2.
Select the required value with the control wheel and press on the control wheel to validate.
To enable the function, press on the
[P0.1]
button. The indicator light comes on and the measurement
begins. The following is displayed in the alarm banner: ‘P0.1 measurement enabled’.
Note:
Pressing on the
[P0.1]
button again disables the function.
Note:
P0.1 measurement is accessible in all modes in invasive ventilation except CPAP and
Duo-Levels. It is not available in non-invasive ventilation, where the risk of leakage does not
guarantee a reliable measurement.
Note:
The set periodicity remains in the memory when the function is disabled.
5.5 NIF (negative inspiratory force)
The NIF measurement can be useful in determining the possibility of a successful extubation.