The enrichment of the O2 mix then takes place according to the FiO2 parameter, and the device then
delivers, as far as possible, the desired concentration. If the ventilation and low-pressure source set-
tings are incompatible, the FiO2 will not be reached and the “low FiO2” alarm will be generated. The
user will then be informed that these settings need to be modified.
CAUTION:
To have efficient operation on a low O2 pressure source, the high pressure O2
connection hose must be disconnected from the device.
To prevent the low pressure O2 fitting becoming detached, ensure the pressure applied to the low
pressure oxygen inlet is less than 150 kPa (1.5 bar).
6.2.4 Sensors
CO2 monitoring
The
[CO2Monitoring]
key activates and deactivates CO2 monitoring.
See CO2 Option on page 43 for more information.
Calibration
The
[CO2 sensor calibration]
key calibrates the CO2 probe.
See CO2 Option on page 43 for more information.