6 Menu
Note:
The suction time is limited to 40 seconds. Afterward, the end-of-aspiration oxygenation
phase (N°3) is automatically started.
Note:
The user can also restart the ordinary ventilation by pressing the
[Validate]
button in
the dialogue box.
3- End-of-suction oxygenation
This step is used to re-oxygenate the patient after his or her disconnection during pulmonary suction.
A countdown indicates the time remaining for oxygenation.
At the end of this countdown, oxygenation shuts down and the
[xx% O2
↗]
key turns off.
Note:
It is possible to shut down the function at any time by pressing the
[xx% O2
↗]
key.
Note:
Throughout the entire oxygenation-aspiration function, nebulization is suspended. It
automatically resumes at the end of the function.
Settings
The different oxygenation settings can be adjusted via a window of the
[O2
↗
Settings]
[Menu]
.
This menu is used to set:
•
Target O2: FiO2 applied to the whole oxygenation-suction period.
•
Duration of oxygenation before suction (step 1 “Oxygenation”)
•
Duration of oxygenation after suction (step 3 “End-of-aspiration oxygenation”)
For more information, refer to Checklist on page 135.
6.3.2 Screen lock key (lock)
Press this button to lock the screen.
To unlock the screen, press this key again, and then validate using the control wheel.
6.3.3 Inhibiting the spirometry sensor
Whenever alarm #57
“Expiratory flow measurement inoperative”
triggers, it appears as a high
priority (red color code).
The user has the ability to inhibit this alarm: