9 Maintenance
CAUTION:
The Monnal Clean’in filter may not be disinfected, cleaned, or sterilized.
Replacing the filter
Remove the filter assembly from its housing with the help of the drawer and replace it.
9.5 Air intake filter
The air intake filter must be managed and cleaned every three months.
To access to the filter, remove the rear panel.
Two cleaning methods are recommended:
•
Remove the filter from its housing, and wash it using warm soapy water. Rinse under running wa-
ter. Make sure that the filter is fully dry before reinstalling it.
•
Remove the filter from its housing and use an air jet to remove any particles trapped in it, and then
reinstall the filter.
9.6 Bacteriological filter
Apply manufacturer recommendations concerning the frequency of replacement of the bacteriological
filter. Refer to the user instructions supplied with the filter.
95
Summary of Contents for MONNAL T75
Page 2: ......