17
1 . Remove the Camera Handpiece from the holder .
2 . Insert the Camera Handpiece tip between the second and third layers
(between the outer paper and plastic layers) with the camera/light source
facing towards the paper layer . See A .
3 . Push the Camera Handpiece completely in until the tip is fully inserted into
the narrow tip of the sheath . See B .
4 . Gently squeeze the illuminated tip of the Camera Handpiece between the
thumb and index finger to ensure that the optically clear area of the cam-
era cover is flat against the lens tip . See C .
5 . Peel away and discard the top plastic layer and the bottom paper layer .
See D .
NOTE:
For optimum image clarity, the optically clear section of the sheath must be aligned with the camera
lens by facing the camera/light source section of the handpiece down toward the paper layer.
Disposable Camera Cover Installation -
Install a new camera cover
on the Camera Handpiece for each patient as follows
:
PRE-OPERATION
2
The camera cover must be used for only one patient and disposed
of properly in accordance with local code.
Camera Cover Installation
POLARIS
CAMERA
HANDPIECE
LENS
CAMERA
COVER
SQUEEZE
PLASTIC
LAYER 2
PLASTIC
LAYER 1
PAPER
LAYER 4
PAPER
LAYER 3
B
A
D
C
The Polaris System should be used with a Camera Cover covering the handpiece.
Make sure to use a new disposable Camera Cover for each patient.