Electromagnetic compatibility (EMC)
Interference immunity measurements on supply input
Immunity to interference, line-conducted disturbances
induced by high-frequency fields – AC voltage grid
IEC 61000-4-6:2013
3 V
0.15 - 80 MHz
6 V
ISM frequency bands
0.15 - 80 MHz
80 % AM at 1 kHz
Conforms
Immunity to interference due to voltage dips, short inter-
ruptions and voltage fluctuations
IEC 61000-4-11:2004
Conforms
Electromagnetic compatibility (EMC)
Interference immunity measurements SIP/SOP
Immunity to interference by discharge of static electricity
IEC 61000-4-2:2008
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Conforms
Immunity to electrical fast transients/bursts – I/O,
SIP/SOP ports
IEC 61000-4-4:2012
± 1 kV
100 kHz repetition frequency
Conforms
Immunity to conducted disturbances, induced by radio-
frequency fields – SIP/SOP ports
IEC 61000-4-6:2013
3 V
0.15 - 80 MHz
6 V
ISM frequency bands
0.15 - 80 MHz
80 % AM at 1 kHz
Conforms
4.2
Phosphor storage plate
Classification
Medical Devices of class
Class IIa
FDA classification (CFR Title 21)
Class I
Ambient conditions during operation
Temperature
°C
18 - 45
°F
64 - 113
Relative humidity
%
< 80
Product description
14
2160100287L41 1911V006
EN-