7.2
Electrical connection
Electrical safety when making connections
❯
Connect the device to a correctly installed
power outlet only.
❯
Do not place non-fixed multi-socket units on
the floor. Comply with the requirements in sec-
tion 16 of IEC 60601-1 (EN 60601-1).
❯
Do not operate any other systems using the
same multiple socket.
❯
Make sure that none of the electrical cables
leading to the unit are under any mechanical
tension.
❯
Before initial start-up verify that the mains sup-
ply voltage and the voltage stated on the type
plate match (see also "4. Technical data").
Connecting the unit to the mains supply
The unit has no main power switch. For
this reason, it is important to set up the
unit properly such that the plug can be
easily accessed and unplugged if
required.
Requirements:
ü
Properly installed power outlet close to the unit
(observe the max. length of the power cord)
ü
Easily accessible power outlet
ü
Mains voltage matches the information shown
on the type plate of the power supply unit
Only the supplied power supply unit may
be used.
❯
Attach the matching country-specific adapter
to the power supply unit.
❯
Remove the screw covers and screws from the
bottom of the device.
❯
Remove the cover from the rear of the device.
Installation
22
2160100287L41 1911V006
EN-
US