DC emitter, 6 mA
Tube length 30 cm
without X-ray field
limitation
X-ray field limitation
2x3
X-ray field limitation
3x4
60 kV
mGycm
2
60 kV
mGycm
2
60 kV
mGycm
2
Incisor
0.10 s
9.1
0.10 s
1.9
0.10 s
3.8
Premolar
0.13 s
11.8
0.13 s
2.5
0.13 s
5.0
Molar
0.16 s
14.8
0.16 s
3.1
0.16 s
6.2
Bitewings
0.16 s
14.8
0.16 s
3.1
0.16 s
6.2
❯
Check and adjust the specific X-ray unit in accordance with the standard values.
8.5
Commissioning tests
The required tests (e.g. acceptance test) must be done in accordance with local rules and regulations.
❯
Find out which tests are required.
❯
Carry out testing in accordance with local rules and regulations.
Acceptance check
The 2D X-ray test phantom is required for acceptance tests with the PSP and sensor as
receivers, and possibly also the corresponding test phantom holder.
❯
Before commissioning the unit, the acceptance test of the X-ray system must be carried out in accor-
dance with national regulations.
The consistency tests, which must be carried out at regular intervals by the dental professional, are
based on the results of the acceptance test.
Electrical safety checks
❯
Carry out the electrical safety check according to national law (e. g. in accordance with IEC 62353).
❯
Document the results.
❯
Carry out and document the instruction and handover for the unit.
A sample handover report is included in the attachment.