WARNING
Risk of cross contamination when not
using the barrier envelope or when
using the barrier envelope more than
once
❯
Do not use an phosphor storage plate
without a barrier envelope.
❯
Do not re-use the barrier envelope
(disposable item).
WARNING
Danger from the re-use of products
intended for single use
Single-use article is damaged after use
and cannot be reused.
❯
Dispose of single-use articles after use.
Preparing the X-ray
ü
The PSP has been cleaned.
ü
The PSP is not damaged.
ü
RFID tag sticks to the phosphor storage plate.
If the RFID tag peels off, replace the phosphor
storage plate.
❯
During first use or after storage for over a
week: erase the PSP (see "10.4 Erasing the
imaging plate").
❯
Completely slide the PSP into the barrier enve-
lope. The white (inactive) side of the PSP must
be visible.
back
IDX
2
*+E247G36120/$+01234567L*
a
back
IDX
2
*+E247G36120/$+01234567L*
a
❯
Pull off the adhesive strip and close the barrier
envelope tightly by pressing together firmly.
back
IDX
2
*+
E247G36120/$+01234567L*
a
back
IDX
2
*+
E247G36120/$+01234567L*
a
1
2
Taking the X-ray image
NOTICE
Damage to the phosphor storage
plate (PSP) caused by a sharp-edged
holding system
❯
Only use holding systems that do not
damage the barrier envelope or the
PSP.
❯
Do not use holding systems with sharp
edges.
Wear hand protection.
❯
Place the PSP in the barrier envelope into the
patient's mouth.
In doing this, make sure that the active side of
the phosphor storage plate faces the X-ray
tube.
back
IDX
2
*+E247G36120/$+01234567L
*
a
❯
Set the exposure time and setting values on
the X-ray unit (see "8.4 X-ray unit settings").
❯
Record the X-ray image.
The image data must be scanned within 30
minutes.
Usage
32
2160100287L41 1911V006
EN-
US