SOLR
®
100% OXYGEN MASK LEVEL 3
TM0063-2
60410X
5-7
13.
Once the anti-suffocation cover is removed, the anti-suffocation spring and anti-suffocation plate will
now be released. Remove them by hand and set to the side
14.
Cup hand around the regulator and using high purity air or nitrogen, blow into the inlet quick disconnect
port to eject the anti-suffocation piston.
15.
Remove O-ring ID .140 from anti-suffocation piston and discard.
Figure 5-18 Using Compressed Air Remove Piston
16.
Using the 5/64 hex driver, remove the two screws from the insert body and discard.
Figure 5-19 Remove Hex Screws from Insert Body
NOTE
SLIGHT PRESSURE MAY BE NEEDED TO FREE THE LEVER
17.
Using a hollow tip screwdriver, turn the crown clockwise while holding the regulator on its side to free
the lever.
Figure 5-20 Turn Crown Clockwise to Free Lever