SOLR
®
100% OXYGEN MASK LEVEL 3
TM0063-2
60410X
5-15
SECTION III LEVEL 3 ASSEMBLY
SOLR
®
100% OXYGEN MASK ASSEMBLY
Before assembly of the SOLR
®
mask, ensure all hard parts have been cleaned for oxygen service. Ensure
you have a clean and cleared work space located in a controlled area. Wipe all work surfaces down with
Isopropyl Alcohol. Nitrile or latex gloves are recommended. All parts cleaned for oxygen service must be
stored in a covered container or re-sealable poly bag until the assembly step calls for the part. Any part that
shows contamination or has become contaminated must be re-cleaned.
Exhaust Assembly
1.
Place the exhaust valve into the exhaust assembly and insert the stem through the hole in the center.
Cut off excess stem.
Figure 5-31 Cut Off Excess Stem
2.
Ensure the exhaust valve is properly seated.
Figure 5-32 Ensure Exhaust Valve is Properly Seated