Page 2: ...AC DS 03 Automated Deactivation System Instruction Manual ...

Page 3: ...03 Printed in the United States of America No part of this publication may be reproduced without the express written permission of AirClean Systems AirClean Systems 2179 E Lyon Station Road Creedmoor NC 27522 USA Tel 800 849 0472 919 255 3220 Fax 919 255 6120 Email info aircleansystems com Web www aircleansystems com ...

Page 4: ...Ratings 1 Warning Symbols 2 Setting up your AC DS 03 2 Securing your AC DS 03 3 Plumbing Installation 4 Electrical Requirements 4 Filter Installation and Maintenance 6 Use of AC DS 03 Automated Deactivation System 6 Cleaning 8 Monthly Cleaning 8 Quick Trouble Shoot 8 Main Wiring Diagram 8 Limited Warranty 9 Specifications and Technical Data 9 ...

Page 5: ... 03 can be used in conjunction with automatic processing devices for endoscopic ultrasound probes Safety Precautions Disconnect power before replacing or servicing the unit Use PPE whenever operating the AC DS 03 wear safety goggles lab coat proper gloves and face eye shield Environmental Conditions For indoor use only Temperature range 5 C to 40 C Main supply voltage fluctuations not to exceed 10...

Page 6: ...AL EQUIPMENT DAMAGE Protective Grounding Setting up your AC DS 03 Included with your new AC DS 03 system is a packet of the following parts Power cord Two E Z mount wall anchors and screws Two wall mounting brackets 10 feet of 3 8 O D drain hose 10 feet of 3 8 O D overflow hose Two Right angle 3 8 barbed adaptors Two Straight 3 8 barbed adaptors One port plug One sample packet of NeutraHyde Part A...

Page 7: ... 03 and install screws loosely 4 Position unit against wall for wall bracket installation Push bracket against wall and mark position of bracket hole See Figure 1 5 Repeat on other side of unit 6 Install the two self drilling wall anchors at these points until they are flush with the wall Place bracket over anchor and install Phillips head screws do not over tighten Now tighten 7 For masonry walls...

Page 8: ...ght side of the AC DS 03 marked INLET As above route the feed line behind or under unit if required 6 The 3 8 overflow tube can be connected to either side Use the 3 8 straight barbed connectors to attach the overflow tubing to the AC DS 03 OVERFLOW connector Make sure that the drain is lower than the outlets for the OVERFLOW on the AC DS 03 There are overflow connectors on both sides of the unit ...

Page 9: ...tomated Deactivation System Manual Ver A03 5 Figure 3 3 1 Typical installations of the AC DS 03 PLEASE CALL AIRCLEAN SYSTEMS AT 1 800 849 0472 WITH ANY INSTALLATION QUESTIONS YOU MAY HAVE Inlet Overflow Discharge ...

Page 10: ...m is designed to be used for neutralizing high level disinfectants before disposal CAUTION Always use proper PPE whenever operating the AC DS 03 1 Once the preceding steps including filter installation have been followed the unit is ready to be plugged in 2 Plug the power cord into the power receptacle on the left side of unit 3 Plug power cord into hospital grade 20 Amp GFCI protected wall recept...

Page 11: ...ng 11 Once tank is empty the ADD NEUTRALIZER LED will flash and operator will repeat cycle starting at step 5 above 12 If tank needs to be emptied any time before reaching the full level the OPTIONAL DISCHARGE button can be pushed at any time to pump out tank contents 13 Unit will then return to step 5 to initiate cycle Figure 4 LED reset and discharge buttons Figure 5 Access cover on top of unit ...

Page 12: ... not running check the following 1 Ensure that the unit is plugged into a live wall outlet 2 Make sure the On Off switch located front panel is on 3 In the event that the system is still not working check the two fuses The fuses are located on the left side of the AC DS 03 in a pull out drawer on the power receptacle CAUTION UNPLUG UNIT BEFORE PULLING OPEN FUSE DRAWER 4 Pull open drawer and remove...

Page 13: ...es obligations or other liabilities expressed or implied What is not covered by this warranty Conditions and damages resulting from any of the following Improper installation delivery within the facility or maintenance Any repair modification alteration or adjustment not authorized by AirClean Systems Misuse abuse accidents or unreasonable use Incorrect electric current voltage or supply Failure t...