15
The following operating instructions are specific for using powders and particulate in your
AirClean
®
Systems PowderSafe
™
enclosure.
NOTE:
Ensure that the INSTALLATION AND ASSEMBLY INSTRUCTION steps have
been completed.
A. To operate the PowderSafe
™
enclosure:
1. Connect the supplied power cord to the unit and to the power outlet. Flip the switch on
the power inlet module. This will activate the main power. The screen on the controller
should light up and display the time and the date. Press power button. Check the blower
to assure it is on . If not, press blower button. Allow the enclosure to run for 5 minutes
before beginning your application.
2. Before use, ensure that you are not using an incompatible substance within the
enclosure. With a HEPA filter,
use only powders and particulate
in the enclosure. For
other substances, please call AirClean
®
Systems to verify compatibility before use (in
this case, your filter may have another model number). Dial 919-255-3220 or
800-849-0472, Monday-Friday 8:00 AM-5:00 PM EST.
3. Perform any part of your process that involves powders and particulate inside your
enclosure. NEVER manipulate powders or particulate at the front edge of your
enclosure.
B. When working within the enclosure, keep the front panels down and preform your
procedure in the center to rear of the enclosure. Never open the front panels completely.
Only open during operation, cleaning or maintenance.
C. If the low airflow alarm sounds and/or lights up, the prefilter or HEPA filter may be clogged.
D. When this happens refer to the “MAINTENANCE” section of this manual.
REMEMBER:
Following the instruction assures better protection of the operator and the
environment, and extends the life of the enclosure and filter. Correct use of the PowderSafe
™
enclosure is the responsibility of the operator!
OPERATING INSTRUCTIONS