background image

Fingertip 

 

Pulse Oximeter

 

USER MANUAL 

Ver1.0C11 

General Description 

Oxygen Saturation is a percentage of Oxyhemoglobin (HbO

2

) capacity, compounded with oxygen, by all combinative hemoglobin 

(Hb)  capacity  in  blood.  In  other  words,  it  is  consistency  of  Oxyhemoglobin  in  blood.  It  is  a  very  important  parameter  for  the 
Respiratory Circulation System. Therefore, it is very important to know the oxygen saturation. 

The  fingertip  pulse  oximeter  features  low  power  consumption,  convenient  operation  and  portability.  Place  one  fingertip  into  the 
photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display.   

Measurement Principle 

Principle  of  the  oximeter  is  as  follows:  A  mathematical  formula  is  established  making  use  of  Lambert  Beer  Law  according  to 
Spectrum Absorption Characteristics of Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. 
Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity 
Pulse  Scanning  and  Recording  Technology,  so  that  two  beams  of  different  wavelength  of  lights  (660nm  glow  and  940nm  near 
infrared  light)  can  be  focused  onto  a  human  nail  tip  through  a  clamping  finger-type  sensor.  A  measured  signal  obtained  by  a 
photosensitive element, will be shown on the oximeter

’s display through process in electronic circuits and microprocessor. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 
2. Red and Infrared-ray Receipt Tube 

Precautions For Use 

Before use, carefully read the manual. 

Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU). 

The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO

2

 measurement. Verify 

that nothing is hindering the pulse measurement before relying on the SpO

2

 measurement. 

Do not use the fingertip pulse oximeter in an MRI or CT environment. 

Do  not  use  the  fingertip  pulse  oximeter  in  situations  where  alarms  are  required.  The  device  has  no  alarms.  It  is  not  for 
continuous monitoring. 

Do not use the fingertip pulse oximeter in an explosive atmosphere. 

The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other 
methods of assessing clinical signs and symptoms. 

Check the pulse oximeter sensor application site every 4 hours to determine the positioning of the sensor and circulation 
and skin sensitivity of the patient. 

Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not 
intended for sterilization. 

10 

Follow  local  ordinances  and  recycling  instructions  regarding  disposal  or  recycling  of  the  device  and  device  components, 
including batteries. 

11 

This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility . However, because of the proliferation 
of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is 
possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of 
this device. 

12 

Portable and mobile RF communications equipment can affect the equipment. 

Inaccurate measurements may be caused by 

Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin); 

Intravascular dyes such as indocyanine green or methylene blue; 

High ambient light. Shield the sensor area if necessary; 

Excessive user movement; 

High-frequency electrosurgical interference and defibrillators; 

Venous pulsations; 

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line; 

The user has hypotension, severe vasoconstriction, severe anemia, or hypothermia; 

The user is in cardiac arrest or is in shock; 

10 

Fingernail polish or false fingernails; 

11 

Weak pulse quality (low perfusion); 

12 

Low hemoglobin; 

Product Properties 

Operation of the product is simple and convenient. 

The product is small in volume, light in weight and convenient to carry. 

Power consumption of the product is low and the two AAA batteries can be operated continuously for 30 hours.   

A  low  voltage  warning  will  be  indicated  when  battery  voltage  is  low  and  normal  operation  of  the  oximeter  might  be 
influenced. 

The product will automatically power off when there is no signal for longer than 8 seconds. 

Intended Use 

Fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin 
(SpO

2

) and pulse rate. 

Operation Instructions 

Install two AAA batteries according to the Battery Installation instructions listed above in the right column. 

Open the clamp as illustrated in the picture below. 

Fully insert one fingertip into the silicone hole of the oximeter before releasing the clamp. 

Press the switch button once on front panel. 

Keep your finger still during measurement. 

Read corresponding data from display screen. 

Front Panel 

The pulse bar graph displays corresponds with the user

’s pulse beat. The height of the bar graph shows the user’s pulse strength. 

 

Product Accessories 

1. 

One lanyard 

2. 

Two AAA batteries 

3. 

One user

’s manual

 

Battery Installation 

1. 

Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) 
signs in the compartment. If the polarities are not matched, damage may be caused to 
the oximeter. 

2. 

Slide the battery door cover horizontally along the arrow shown as the picture. 

Notes:   

 

Install the batteries with the correct polarity. Incorrect placement may cause damage to 
the bracket. 

 

Please remove the batteries if the pulse oximeter will not be used for long periods of time. 

Using the Lanyard 

1. 

Thread thinner end of the lanyard through the loop. 

2. 

Thread thicker end of the lanyard through the threaded end before pulling it tightly. 

Warnings! 

 

Keep the oximeter away from young children. Small items such as the battery door, 
battery, and lanyard are choking hazards. 

 

Do not hang the lanyard from the 

device’s electrical wire. 

Maintenance and Storage 

1. 

Replace the batteries in a timely manner when low voltage lamp is lighted. 

2. 

Clean surface of the fingertip oximeter before it is used. 

3. 

Remove the batteries if the oximeter is not operated for a long time. 

4. 

It is best to store the product in -20

℃~+

55

 

and 

93% humidity

.

 

5. 

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage. 

6. 

Dispose of battery properly; follow any applicable local battery disposal laws. 

The  use  life  of  the  device  is  five  years  when  it  is  used  for  15  measurements  every  day  and  10  minutes  per  one 
measurement. Stop using and contact local service center if one of the following cases occurs: 

 

An error in the 

Possible Problems and solutions 

is displayed on screen. 

 

The oximeter cannot be powered on in any case and not the reasons of battery. 

 

There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or the 

key is unresponsive or unavailable. 

Specifications 

1. Display Type 

LED display 

2. SpO

2

 

Display range: 0-99% 
Measurement range: 70-99%   
Accuracy: 70%-99%: ±3%;    0%~69% no definition 
Resolution: 1%   

3. Pulse Rate 

Display range: 0~254BPM 
Measure range: 30-235 BPM 
Accuracy: 30~99bpm, ±2bpm;    100~235bpm, ±2% 
Resolution: 1BPM 

4. Probe LED Specifications 

 

Wavelength 

Radiant Power 

RED 

660

±

2nm 

1.8mW 

IR 

940

±

10nm 

2.0mW 

5. Power Requirements 

Two AAA alkaline Batteries 
Power consumption: Less than 25mA   

Low power indication:

 

 

Battery Life: Two AAA 1.5V, 600mAh alkaline batteries could be continuously operated as long as 30 hours.   
It is equipped with a function switch, through which the oximeter can be powered off in case no finger is the oximeter longer than 8 
seconds. 

6. Outline Dimension 

Length: 56mm 
Width: 32mm 
Height: 30mm 
Weight: 50g (including two AAA batteries) 

7. Environment Requirements 

Operation Temperature: 5

40

 

Storage Temperature: -20

~+

55

 

Ambient Humidity: 15%~80% in operation; 

93% in storage 

8. Equipment Response Time 

As shown in the following figure. 
Response time of slower average is 12.4s. 

 

9. Classification 

According to the type of protection against electric shock: INTERNALLY POWERED EQUIPMENT; 
According to the degree of protection against electric shock: TYPE BF APPLIED PART;

 

According  to  the  method(s)  of  sterilization  or  disinfection  recommended  by  the  manufacturer:  Equipment  with  method(s)  of 
sterilization or disinfection recommended by the manufacturer; 
According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH 
OXYGEN OR NITROUS OXIDE: EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTURE. 
According to the degree of protection against ingress of water: IPX1 
According to the mode of operation: CONTINUOUS OPERATION 
 
 
 

Reviews: