Troubleshooting

1. After the machine is turned on, the low water indicator lights up, while the water tank is

already full, and the machine does not start heating. This may be caused by the bad contact of
the water level sensor which is caused by the rust from the water. To solve the problem,
unplug and open the machine, find the two wires which connects to the power cord at the top
of the boiler, remove the power cord, loosen the stainless steel stick and remove from the top,
wipe the stick clean and put it back into the machine.

2. The machine start heating after the power is on, and the pressure continue to rise after 4.5MPa.

This may be caused by: 1. Relay (needed for replacement). 2. Pressure regulator (needed for
replacement).