Data Logger Manual
Section 1: Device Safety
CAUTION:
DEVICE MUST BE USED ONLY IN A MANNER CONSISTENT WITH THIS MANUAL.
NOTICE:
WHEN 230VAC SUPPLY USED, OMEGA SPECIFIES USE OF AC SUPPLY PART # T35-9-100R-3
MANUFACTURED BY ENG ELECTRIC. THIS SUPPLY IS AVAILABLE FROM OMEGA.
1.1 Model information
• Model:
OM-CP-PR2000
•
Description: Pressure Logger with LCD display
•
Manufactured in the USA
1.2 Specifi cations
Pressure
Range: 0 to 30, 100, 300 and 500 PSIA/G, 0 to 1000 and 5000PSIA
Resolution: 0.002, 0.005, 0.02, 0.05PSIA/G
Calibrated Accuracy: 2% FSR, 0.25% @ 25°C typical
Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text:
Confi gurable channel text size
Indicators: Power, status, memory
Backlight:
Confi gurable w/auto shut-off and contrast adjustment
Start/Stop Time:
Software programmable start time and date, up to six months in advance; programmable stop time
Memory:
262,143
total
readings; software confi gurable memory wrap
Reading Rate:
1 reading every 2 seconds to 1 every 24 hours
Calibration:
Digital calibration through software
Calibration Date:
Automatically recorded within device
Battery Type:
6 alkaline AA batteries, user replaceable; optional AC adapter, 7-24 VDC, 100mA. For 230 VAC operation use AC
supply T35-9-100 R-3 made by ENG Electric.
Battery Life:
60 days typical with continuous screen use; 360 days with display off.
Data Format:
Date and time stamped PSI, inHg, mmHg, bar, atm, Torr, Pa, kPa, MPa
Time Accuracy:
±1 minute/month (at 20º to 30°C)
Computer Interface:
PC serial or USB (interface cable required); 115,200 baud
Start/Stop Time:
Software programmable start time and date, up to six months in advance; programmable stop time
Reading Rate:
1 reading every 2 seconds to 1 every 24 hours
Battery Life:
60 days typical with continuous screen use; 360 days with display off.
Page 3
Calibration Date:
Automatically recorded within device