background image

 

HOBO Temp Data Logger (UX100-001) Manual 

 

3.

 

Select the Low Level checkbox if you want to set up a 
condition in which burst logging will occur when the sensor 
reading falls below the low limit value. Type in the value or 
drag the blue lower slider. 

4.

 

Set the burst logging interval, which must be less than the 
logging interval. Select either a preset burst logging interval 
or select Custom and enter your own interval. Keep in mind 
that the more frequent the burst logging rate, the greater 
the impact on battery life and the shorter the logging 
duration. 

5.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging Mode 
in the Launch Logger window to make additional changes. 

Notes: 

 

Once the logger is launched, the high and low burst 
logging levels are only checked when the logger's 
LCD screen refreshes once every 15 seconds. Therefore, 
if you set the logging interval to less than 15 seconds and 
the sensor reading falls outside the levels, the burst 
logging will not begin until the next 15-second refresh 
cycle. 

 

The actual values for the burst logging levels are set to 
the closest value supported by the logger. For example, 
the closest value to 85°F that the logger can record is 
84.990°F and the closest value to 32°F is 32.043°F. 

 

Burst logging mode can begin or end when the sensor 
reading is within the logger specifications of 0.02°C 
resolution. This means the value that triggers burst 
logging may differ slightly than the value entered. For 
example, if the High Level for a temperature alarm is set 
to 75.999°F, burst logging can start when the sensor 
reading is 75.994°F (which is within the 0.02°C 
resolution). 

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded in “normal mode.” For example, let’s assume 
the logger has a 10-minute logging interval and logged a 
data point at 9:05. Then, the high level was surpassed 
and burst logging began at 9:06. Burst logging then 
continued until 9:12 when the sensor reading fell back 
below the high level. Now back in normal mode, the next 

logging interval will be 10 minutes from the last burst 
logging point, or 9:22 in this case. If burst logging had not 
occurred, the next data point would have been at 9:15. 

 

A New Interval event will appear on the plot (if you select 
events for plotting in the Plot Setup window) each time 
the logger enters or exits burst logging mode. 

Statistics 

Statistics is a logging mode in which the logger calculates 
maximum, minimum, average, and standard deviation statistics 
during logging, recording the results at each logging interval 
based on samples taken at a rate you specify. This will result in 
up to four additional series per sensor that record the following 
information at each logging interval: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 
values. 

For example, let’s say the logging interval is set to 5 minutes 
and the sampling interval is set to 30 seconds (with maximum, 
minimum, average, and standard deviation all enabled). Once 
logging begins, the logger will measure and record the actual 
temperature sensor values every 5 minutes. In addition, the 
logger will take a temperature sample every 30 seconds and 
temporarily store it in memory. The logger will then calculate 
the maximum, minimum, average, and standard deviation using 
the samples gathered over the previous 5-minute period and 
log the resulting value(s). When reading out the logger, this 
would result in 5 data series: one temperature series (with data 
logged every 5 minutes) plus four maximum, minimum, 
average, and standard deviation series (with values calculated 
and logged every 5 minutes based on the 30-second sampling). 

To set up statistics: 

1.

 

Select Statistics for Logging Mode in the Launch Logger 
window. If Statistics has already been configured for this 
logger, click the Edit button in the Launch Logger window. 

2.

 

Click the Maximum, Minimum, Average, and Standard 
Deviation checkboxes for each of the statistics you want to 
calculate during logging. Note that Average is automatically 
enabled when selecting Standard Deviation. In addition, the 
more statistics you record, the shorter the logger duration 
and the more memory is required. 

 

3.

 

Set the sampling interval, which must be less than and a 
factor of the logging interval. Choose either a preset 
sampling interval or select Custom and enter your own 
sampling interval. Keep in mind that the more frequent the 
sampling rate, the greater the impact on battery life. 

Summary of Contents for Onset HOBO UX100-001

Page 1: ...r statistics Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Modes When memory full push button or date time Restart Mode Push button Time Accuracy 1 minute per month at 25 C 77 F see Plot B Battery Life 1 year typical with logging rate of 1 minute and sampling interval of 15 seconds or greater Battery Type One 3V CR2032 lithium batte...

Page 2: ...itely with newest data overwriting the oldest data When first launched all five segments in the memory bar will be empty In this example the memory is full all five segments are filled in and new data is now overwriting the oldest data This will continue until the logger is stopped or the battery runs out The logger is currently logging A sensor reading is above or below the high or low alarm that...

Page 3: ... In Normal mode data will always be recorded at the regular logging interval set in the previous step This is the default setting Burst In Burst mode logging will occur at a different interval when a specified condition is met See Burst Logging for more information Statistics In Statistics mode maximum minimum average and standard deviation statistics are calculated for the temperature during logg...

Page 4: ...y on the LCD screen Note that the alarm limits are only checked when the logger s LCD screen refreshes every 15 seconds The actual values for the high and low alarm limits are set to the closest value supported by the logger For example the closest value to 85 F that the UX100 series logger can record is 84 990 F and the closest value to 32 F is 32 043 F In addition alarms can trip or clear when t...

Page 5: ...or 9 22 in this case If burst logging had not occurred the next data point would have been at 9 15 A New Interval event will appear on the plot if you select events for plotting in the Plot Setup window each time the logger enters or exits burst logging mode Statistics Statistics is a logging mode in which the logger calculates maximum minimum average and standard deviation statistics during loggi...

Page 6: ...to mount it a wall or other flat surface Use the double sided tape to affix the logger to a surface Insert the hook and loop strap through the mounting loops on both sides of the logger to mount it to a curved surface such as a pipe or tubing Protecting the Logger The logger is designed for indoor use and can be permanently damaged by corrosion if it gets wet Protect it from condensation If the me...

Page 7: ...HOBOware provides the option of recording the current battery voltage at each logging interval which is disabled by default Recording battery life at each logging interval takes up memory and therefore reduces logging duration It is recommended you only record battery voltage for diagnostic purposes Even with the channel disabled a bad battery event will still be recorded Office Jl Radin Inten II ...

Reviews: