ATS9360 User Manual
47
Timebase System
Timebase options
Internal Clock or
External Clock (Optional)
Internal Sample Rates
1.8 GS/s, 1.5 GS/s, 1.2 GS/s,
1 GS/s, 800 MS/s, 500 MS/s, 200 MS/s,
100 MS/s, 50 MS/s, 20 MS/s, 10 MS/s,
5 MS/s, 2 MS/s, 1 MS/s, 500 KS/s,
200 KS/s, 100 KS/s, 50 KS/s,
20 KS/s, 10 KS/s, 5 KS/s, 2 KS/s, 1 KS/s
Internal Clock accuracy
±2 ppm
Dynamic Parameters
Typical values measured on the 200 mV range of CH A of a randomly
selected ATS9360. Input signal was provided by a Marconi 2018A signal
generator, followed by a 9-pole, 10 MHz band-pass filter (TTE Q36T-10M-
1M-50-720BMF). Input frequency was set at 9.9 MHz and output amplitude
was 135 mV rms, which was approximately 95% of the full scale input.
Input was averaged
.
SNR 57.1
dB
SINAD 56.6
dB
Note that these dynamic parameters may vary from one unit to another, with
input frequency and with the full scale input range selected.