48
ATS9360 User Manual
Optional ECLK (External Clock) Input
Signal Level
±200mV Sine wave or square wave minimum.
±1V
max.
Input impedance
50
Input Coupling
AC
Maximum frequency
1.8 GHz for Fast External Clock
10 MHz Clock Reference:
10.1 MHz
Minimum frequency
300 MHz for Fast External Clock
10 MHz Clock Reference:
9.9 MHz
Sampling Edge
Rising
Sample Rates Available With
10 MHz Clock Reference:
1.8 GS/s
Note that the accuracy and stability of these
sampling frequencies is dependent on the
accuracy and stability of the 10 MHz Clock
Referece input supplied by the user
Triggering System
Mode
Edge triggering with hysteresis
Number of Trigger Engines
2
Trigger Engine Combination OR
Trigger Engine Source
CH A, CH B, EXT, Software or None,
independently software selectable for each of
the two Trigger Engines
Hysteresis
±5% of full-scale input, typical
Trigger sensitivity
±10% of full scale input range.
This implies that the trigger system may not
trigger reliably if the input has an amplitude
less than ±10% of full-scale input range
selected
Trigger level accuracy
±5%, typical, of full-scale input range of the
selected trigger source