Electromagnetic immunity
41
Immuni-
ty tests
IEC
60601
test
level
Com-
pliance
level
Electromagnet-
ic environment
– guidelines
Voltage
drops,
short
inter-
ruptions
and fluc-
tuations
in the
supply
voltage
in accor-
dance
with
IEC
61000-
4-11
< 5 %
U
T
(> 95
% drop
of U
T
)
for ½
period
40 % U
T
(60 %
drop of
the U
T
)
for 5 pe-
riods 70
% U
T
(30
% drop
of U
T
)
for 25
periods
< 5 % U
T
(> 95 %
drop in
U
T
) for
5 s
Not ap-
plicable
The quality of
the supply volt-
age should cor-
respond with a
typical store or
hospital envi-
ronment. If the
user of the DSO
364 model also
requires that
the function
continues after
interruptions
to the power
supply, it is
recommended
that you supply
the DSO 364
model from an
uninterrupted
power supply
or battery.
Magnetic
field with
supply
frequen-
cy (50/60
Hz) in ac-
cordance
with IEC
61000-
4-8
3 A/m
3 A/m
Magnetic fields
associated with
the mains fre-
quency should
correspond
to the typical
levels encoun-
tered in a store
or hospital
environment.
NOTICE
: U
T
is the mains AC voltage prior to appli-
cation of the test levels.