Electromagnetic immunity
45
b
. Not applicable over a frequency range from
150 kHz to 80 MHz.
Recommended safety clearance between porta-
ble and mobile HF telecommunications devices
and the [DEVICE or SYSTEM].
The DSO 364 model is intended for operation in
an electromagnetic environment where HF dis-
turbances are monitored. The customer or user
of the DSO 364 model can help to avoid electro-
magnetic disturbances by observing the mini-
mum clearance between portable and mobile HF
telecommunication devices (transmitters) and
the DSO 364 model – depending on the output
power of the communication device as discussed
below.
Nominal
output
of the W
transmit-
ter
Safety clearance depending on the
transmission frequency m
150 kHz to
80 MHz
d=1.2 √P
80 MHz to
800 MHz
d=1.2 √P
800 MHz
to
2.5 GHz
d=2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters whose nominal output is not
specified in the above table, the distance can be
determined using the equation in the respective
column where P represents the nominal output
of the transmitter in watts (W) as specified by the
transmitter manufacturer.