65
Table 5 – Signal input/output parts Port
Phenomenon
Basic EMC standard
Electrical fast
transients/burst
IEC 61000-4-4
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
Immunity test levels
Professional healthcare facility environment
±1 kV
100kHz repetition frequency
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and 80MHz 80%AM at 1kHz