NOTE: ATTEMPTED REPAIR, MODIFICATIONS, OR TAMPERING BY UNAUTHORIZED PERSONNEL WILL
VOID THE WARRANTY.
6.3
If the ERROR#2 shows on LCD
If the ERROR#2 shows on LCD, the calibration data in the EEPROM has been erased. The power supply needs to be recalibrat-
ed. Please process the EEPROM reset procedure below and follow the calibration procedures in the manual and enter the cor-
rect calibration value, the ERROR#2 will disappear as well as the beep sound.
EEPROM Reset Procedure for LPS 305:
Press the -Iset(5) key and the +
▼
key simultaneously.
6.4 Using the RS-232-C Serial Interface
This section describes how hoe to set up the RS-232-C interface for remote control.
The interface of LPS305 are designed in accordance with EIA (Electronic Industries Association)
standard RS-232-C. Through its interface be LPS305TPT-3025 be remotely controlled and transmit its internal data to a hose
computer.
Command
Description
Example
VSET1
CH1 Voltage setup
VSET1 5
VSET2
CH2 Voltage setup
VSET2 10
ISET1
CH1 Current setup
ISET1 1.5
ISET2
CH2 Current setup
ISET2 2.5
VOUT1
CH1 Voltage Readback
VOUT1
VOUT2
CH2 Voltage Readback
VOUT2
IOUT1
CH1 Current Readback
IOUT1
IOUT2
CH2 Current Readback
IOUT2
OUT
0=+/- output off
OUTO
1=+/- output on
OUT1
TRACK
0= independent
TRACK0
1= tracking from ch1
TRACK1
2= tracking from ch2
TRACK2
STATUS
Working status(see note7)
STATUS
USER MAINTENANCE/SERVICE