CALI
0=Reset
CALI0
1=begin calibration
CALM
2=input calibration parameter
CALI2 9.574
MODEL
Display model no.
MODEL
VERSION
Display version no.
VERSION
HELP
Display command list
HELP
BEEP
0= beeper function off
BEEP0
1= beeper function on
BEEP1
VDD
0= digital output off
VDD0
3= digital output 3.3V
VDD3
5= digital output 5V
VDD5
NOTE:
1. All RS232 commands are not case sensitive case-nonsensitivity ASCII codes.
2. Use async framing 8 data bits, no parity bit, 1 stop bit.
3. Bit rate=2400bps.
4. Every command string is terminated by CR or LF or BOTH (carriage return)
5. There is only one command allowed in a command string.
6. A command string enters before “OK” prompt will be reject & no function.
7. STATUS Command operation explanation:
The power supply responses the “STATUS” command with a 10 bit binary number. Each bit representing the information as
below.
Bit 0:
channel 1 0=CV 1=CC
Bit 5:
0:digital output 5V
1:digital output 3.3V
Bit 9:
0:beeper function off
1:beeper function on
Bit 1:
channel 2 0=CV 1=CC
Bit 6:
0: +/- output off
1: +/- output on
Bit 10:
0:CC output compensated off
1:CC output compensated on
Bit 3,2:
00:independent
10:tracking to channel 1
11:tracking to channel 2
Bit 7:
0:nothing
1:digital output overload
Bit 4:
0:digital output off
1:digital output on
Bit 8:
0:fan off
1:fan on
USER MAINTENANCE/SERVICE
Summary of Contents for AF Series
Page 1: ...Linear Programmable Power Supply AF2303P User Manual AF SERIES PRODUCT OF ALLFTEK ...
Page 2: ......
Page 10: ...FIGURE A GETTING STARTED ...
Page 14: ...FIGURE C GETTING STARTED ...
Page 23: ......