CALIBRATION
The 9850 T/A is calibrated by first setting the
zero reading mechanically and electrically,
and then setting the low range span and high
range span.
This procedure should only be attempted by
persons who have the proper equipment. Any
attempt at recalibrating voids all calibration
certification provided with the instrument. To
assure best calibration results this procedure
should be performed in standard room
conditions, that is, 25 degrees Celsius and 760
mm Hg barometric pressure.
Note:
A change in any one adjustment
requires calibration of the zero adjustment,
high range span adjustment, and low range
span adjustment.
WARNING
: To maintain the instrument as
intrinsically safe for use in hazardous
locations, use
ONLY
the following battery
types: Eveready 522; Duracell MN1604;
Radio Shack 23-553.
LIST OF MODELS AND PARTS
Part No.
Description
634-210-100
9850 Analog Thermo-
Anemometer
(marked in FPM)
634-210-115
9850 Analog Thermo-
Anemometer
(marked
in
meters/sec)
346-028-003 Soft
Carry
Pouch
459-010-011 Battery,
9V
Alkaline
116-210-001 Owner's
Manual