5 | Checking
BYK-Gardner GmbH
26
wave-scan 3 ROBOTIC
301 200 141 E 2010
5.3 Procedure
Perform a measurement with the robot. The values for a correct reading
are printed in the retraceable certificate and on the reference tile. These
values can be entered in smart-robotic for a pass / fail check.
Illustration 26:
Daily check in smart-robotic
If the value measured on the reference tile is within the printed tolerance
range, the requirements are met.
If the mean value is not within the desired tolerance range, try carefully
cleaning the test tile. If this produces no improvement, please contact our
customer service department.