BYK-Gardner GmbH
Getting Started | 3
302 500 060 E 2207
byko-swing
15
3.3 Inserting the Pendulum
Place the pendulum exactly with the small holes on the tapered pins of the lifting
table.
The marking arrow on the bar of the pendulum
must point to the front
.
CAUTION
If pendulum is placed the wrong way, light barriers and counter are inoperable.
To check the correct position:
1. Close the door of the protective cover.
2. Check the position of the pendulum to the scale. The completely stopped pen-
dulum must point exactly to the zero marking of the scale.
3. If not, check the tip of the pendulum. If twisted, straighten it.
4. Repeat this procedure three times to assure an equal rest.
5. Double check the bubble level, correct the leveling if necessary.
6. When bubble is centered, lock the screw feet with the counter nuts.
The
byko-swing
hardness tester is now ready for operation. You can check the
correct function of your instrument with the procedure described in
.