To perform a measurement:
1. Power on the instrument.
2. Place sample on the automatic lifting table.
3. Clean the balls of the pendulum with isopropanol.
4. Select appropriate method with button
Modus
, status LED lights up blue for
Confirm
.
5. Confirm with
Enter
, status LED lights up green for
Ready
.
6. Optional: Adjust duration of acoustic end signal with button
Signal
and con-
firm with
Enter
.
7. Press button
Start/Stop
, status LED lights up red for
Busy
.
8. Measurement starts, wait until measurement finishes.
9. Keep the door closed during measurement.
At the end of the measurement the LED lights up green and the optional signal
sounds. The measurement result is shown on the display.
NOTICE
1 The device can only be operated with the door closed - the clamping protection
is necessary due to the automation - a message "Door open" appears.
2 If selected measurement mode does not match the inserted pendulum, a mes-
sage "Wrong pendulum" appears.
3 Measured values will always be lower than the results on the glass standard –
see
.