5.3 Troubleshooting
Problem
Cause
Solution
Counter does not work
properly.
Pendulum not attached
properly.
Attach pendulum with ar-
row to the front.
Balls of pendulum are
dirty.
Clean the polished balls
of the pendulum.
Message "Wrong pendu-
lum" appears.
Pendulum does not
match selected modus.
Choose correct modus or
changeover pendulum.
Message "Door open"
appears.
Measurements are possi-
ble only with door closed.
Close the door before
starting.
Message "Abort" ap-
pears.
Measurement was inter-
rupted before finishing.
Do not press start/stop
button before measure-
ment finishes.
Door was opened before
finishing.
Keep door closed during
measurement.
Button (start or power
off) seems not to work.
Changes not confirmed.
LED lights up blue.
Confirm the last change.
LED lights up green. Re-
peat.
Pendulum moves out of
position.
Pendulum not placed on
the pins of the lifting ta-
ble.
Place pendulum exactly
with the small holes on
the pins of the lifting ta-
ble.
No display.
Power supply not
plugged in properly.
Check power connection.